5.1. ABORt

5.1. ABORT

Abort commands cancel any triggered actions.

 

SCPI command

Description

ABORt

Resets the trigger system to the Idle state

:DLOG

Stops the internal data logging session

5.1.1. ABORt

Syntax

ABORt

Description

The ABORt command resets the trigger system and places all trigger sequences in the IDLE state. Any actions related to the trigger system that are in progress will be also aborted as quickly as possible. As a result, subsequent triggers have no effect on the input level.

 

ABORt is also executed at power-on and upon execution of the *RST command.

Usage example

ABOR

Related Commands

*RST

INITiate

[SOURce[<n>]]:CURRent:TRIGgered

[SOURce[<n>]]:VOLTage:TRIGgered

[SOURce[<n>]]:LIST:COUNt

5.1.2. ABORt:DLOG

Syntax

ABORt:DLOG

Description

This command stops the internal data logging session.

 

ABORt:DLOG is also executed at power-on and upon execution of the *RST command.

Usage example

ABOR:DLOG

Related Commands

*RST