8. Parameters and settings
8.1. Programming parameters
The BB3 firmware can be build to support any combination of various type of peripheral modules that that are inserted into BB3 chassis.
Use the *IDN? command to find out what channels are defined in the BB3’s firmware.
8.1.1. Voltage
Power module model |
DCP405 |
DCM220 |
MAXimum [V] |
40 |
20 |
MINimum [V] |
0 |
1 |
DEFault [V] |
0 |
0 |
Value after *RST or *TST? |
0 |
0 |
STEP MINimum [V] |
10m |
10m |
STEP MAXimum [V] |
5 |
2 |
STEP DEFault [V] |
100m |
100m |
PROTection DELay MINimum [s] |
0 |
0 |
PROTection DELay MAXimum [s] |
10 |
10 |
PROTection DELay DEFault [s] |
5m |
5m |
CALibration VALue MINimum[V] |
0.15 |
2 |
CALibration VALue MIDdle[V] |
20 |
10 |
CALibration VALue MAXimum[V] |
38 |
18 |
8.1.2. Current
Programming range or model |
DCP405 |
DCM220 |
MAXimum [A] |
5 |
4 |
MINimum [A] |
0 |
0 |
DEFault [A] |
0 |
0 |
Value after *RST or *TST? |
0 |
0 |
STEP and STEP MINimum [A] |
10m |
10m |
STEP MAXimum [A] |
1 |
1 |
STEP DEFault [A] |
50m |
50m |
PROTection DELay MINimum [s] |
0 |
0 |
PROTection DELay MAXimum [s] |
10 |
10 |
PROTection DELay DEFault [s] |
20m |
20m |
CALibration VALue MINimum[A] |
50m |
500m |
CALibration VALue MIDdle[A] |
2.425 |
2 |
CALibration VALue MAXimum[A] |
4.8 |
3.5 |
8.1.3. Power
The total number of possible combinations for the BB3’s power parameters is 6, of which two typical examples are given below:
Programming range or model |
DCP405 |
DCM220 |
MINimum [W] |
0 |
0 |
DEFault [W] |
155 |
80 |
MAXimum [W] |
155 |
80 |
PROTection LEVel DEFault [W] |
155 |
80 |
PROTection DELay MINimum [s] |
1 |
1 |
PROTection DELay MAXimum [s] |
300 |
300 |
PROTection DELay DEFault [s] |
10 |
10 |
8.2. Reset Settings (*RST)
At power-on or after execution of the *RST common command, device settings will be set to the states listed in the following table. See also the MEMory:STATe:RECall:AUTO command.
Command |
Power on |
*RST |
*ESE |
0 |
Not affected |
*ESR |
0 |
Not affected |
*SRE |
0 |
Not affected |
*STB? |
0 |
??? |
CAL[:MODE] |
OFF |
|
CAL:STAT |
ON if valid calibrating data for both voltage and current exists in the non-volatile memory, otherwise OFF. |
|
DISP:BRIG |
Power down state |
20 |
DISP:VIEW |
Power down state |
1 |
DISP[:STAT] |
Power down state |
Not affected |
INST:NSEL |
1 |
|
INST:SEL |
CH1 |
|
MEMM:CDIR |
/ |
|
MEMM:DOWN:FNAM |
"" |
|
MMEM:LOCK |
Power down state |
Not affected |
OUTP[:STAT] |
Power down state |
OFF |
OUTP:DPR |
Power down state |
Not affected |
OUTP:PROT:COUP |
Power down state |
OFF |
OUTP:TRAC |
Power down state |
OFF |
SENS:CURR:RANG |
Power down state |
BEST |
SENS:DLOG:FUNC:CURR |
Power down state |
OFF |
SENS:DLOG:FUNC:POW |
Power down state |
OFF |
SENS:DLOG:FUNC:VOLT |
Power down state |
OFF |
SENS:DLOG:PER |
Power down state |
0.02 |
SENS:DLOG:TIME |
Power down state |
60 |
[SOUR[n]]:CURR |
Power down state |
DEF (see Section 8.1) |
[SOUR[n]]:CURR:PROT:DEL |
Power down state |
DEF (see Section 8.1) |
[SOUR[n]]:CURR:PROT:STAT |
Power down state |
OFF |
[SOUR[n]]:CURR:PROT:TRIP? |
0 |
|
[SOUR[n]]:CURR:STEP |
Power down state |
DEF (see Section 8.1) |
[SOUR[n]]:POW:PROT[:LEV] |
Power down state |
DEF (see Section 8.1) |
[SOUR[n]]:POW:PROT:DEL |
Power down state |
DEF (see Section 8.1) |
[SOUR[n]]:POW:PROT:STAT |
Power down state |
ON |
[SOUR[n]]:POW:PROT:TRIP? |
0 |
|
[SOUR[n]]:VOLT |
Power down state |
DEF (see Section 8.1) |
[SOUR[<n>]]:VOLT:PROG[:SOUR] |
INT |
|
[SOUR[n]]:VOLT:PROT:DEL |
Power down state |
DEF (see Section 8.1) |
[SOUR[n]]:VOLT:PROT:STAT |
Power down state |
OFF |
[SOUR[n]]:VOLT:PROT:TRIP? |
0 |
|
[SOUR[n]]:VOLT:STEP |
Power down state |
DEF (see Section 8.1) |
[SOUR[<n>]]:VOLT:SENS[:SOUR] |
Power down state |
OFF |
STAT:OPER[:EVEN] |
0 |
Not affected |
STAT:OPER:COND |
0 |
Not affected |
STAT:OPER:ENAB |
0 |
Not affected |
STAT:OPER:INST[:EVEN] |
0 |
Not affected |
STAT:OPER:INST:COND |
0 |
Not affected |
STAT:OPER:INST:ENAB |
0 |
Not affected |
STAT:OPER:INST:ISUM[:EVEN] |
0 |
Not affected |
STAT:OPER:INST:ISUM:COND |
0 |
Not affected |
STAT:OPER:INST:ISUM:ENAB |
0 |
Not affected |
STAT:QUES[:EVEN] |
0 |
Not affected |
STAT:QUES:COND |
0 |
Not affected |
STAT:QUES:ENAB |
0 |
Not affected |
STAT:QUES:INST[:EVEN] |
0 |
Not affected |
STAT:QUES:INST:COND |
0 |
Not affected |
STAT:QUES:INST:ENAB |
0 |
Not affected |
STAT:QUES:INST:ISUM[:EVEN] |
0 |
Not affected |
STAT:QUES:INST:ISUM:COND |
0 |
Not affected |
STAT:OPER:INST:ISUM:ENAB |
0 |
Not affected |
SYST:CHAN:INFO:ONT:LAST? |
0 |
|
SYST:CHAN:INFO:ONT:TOT? |
Power down state |
Not affected |
SYST:CPU:INFO:ONT:LAST? |
0 |
|
SYST:CPU:INFO:ONT:TOT? |
Power down state |
Not affected |
SYST:ERR:COUN? |
0 |
|
SYST:POW |
Power down state |
ON |
SYST:TEMP:PROT [AUX] |
Power down state |
55 |
SYST:TEMP:PROT:DEL [AUX] |
Power down state |
30 |
SYST:TEMP:PROT:STAT [AUX] |
Power down state |
ON |
SYST:TEMP:PROT CH1|CH2 |
Power down state |
75 |
SYST:TEMP:PROT:DEL CH1|CH2 |
Power down state |
30 |
SYST:TEMP:PROT:STAT CH1|CH2 |
Power down state |
ON |
TRIG:DEL |
Power down state |
0 |
TRIG:SLOP |
Power down state |
POS |
TRIG:SOUR |
Power down state |
IMM |
TRIG:SOUR:DLOG |
Power down state |
IMM |
8.3. Special modes of operation
Operations will differ when tracking or any of coupling modes between channels have been selected. The following table shows how special modes of operation affects some features:
|
Calibration (CAL) |
Remote sensing (VOLT:SENS) |
Voltage |
Tracking |
Coupling (INST:COUP:TRAC) |
TRACking |
Disabled |
|
Disabled |
|
Disabled |
Common ground |
Disabled |
|
|
|
|
Split rails coupling |
Disabled |
|
|
|
|
Coupled in PARallel |
Disabled |
|
Disabled |
Disabled |
|
Coupled in SERies |
Disabled |
Disabled |
Disabled |
Disabled |
|
8.4. Default settings
Function |
State / Value |
SCPI command reference |
Over-current protection (OCP) |
OFF |
CURR:PROT:STAT |
Over-voltage protection (OVP) |
OFF |
VOLT:PROT:STAT |
Over-power protection (OPP) |
ON |
POW:PROT:STAT |
OPP trip level |
155.00 W |
POW:PROT:TRIP |
OPP delay |
10 s |
POW:PROT:DEL |
Channel Over-temperature protection (OTP) |
ON |
SYST:TEMP:PROT:STAT |
Channel OTP trip level |
60.00 oC |
SYST:TEMP:PROT:TRIP |
Channel OTP delay |
30 s |
SYST:TEMP:PROT:DEL |
AUX sensor OTP |
ON |
SYST:TEMP:PROT:STAT AUX |
AUX sensor OTP trip level |
50 oC |
SYST:TEMP:PROT:TRIP AUX |
AUX sensor OTP delay |
10 s |
SYST:TEMP:PROT:DEL AUX |
Shutdown when protection tripped |
OFF |
SYST:POW:PROT:TRIP |
Switch off all outputs when protection tripped |
OFF |
OUTP:PROT:COUP |
Remote voltage programming |
OFF |
VOLT:PROG |
Remote voltage sense |
OFF |
VOLT:SENS |
Coupling mode |
NONE |
INST:COUP:TRAC |
Tracking mode |
OFF |
OUTP:TRAC |
Calibration password |
eezbb3 |
CAL:PASS:NEW |
System password |
Not defined |
SYST:PASS:NEW |
Communication mode |
Local |
SYST:LOC |
Front panel lock |
OFF |
SYST:KLOC |
Front panel display |
ON |
DISP:STAT |
Ethernet communication |
OFF |
SYST:COMM:ENAB 0, ETH |
Force disabling of all outputs on power up |
OFF |
SYST:PON:OUTP:DIS |
Beeper |
ON |
SYST:BEEP:STAT |
“Key pressed” click tone |
OFF |
SYST:BEEP:KEY:STAT |
Ethernet port communication |
5025 |
SYST:COMM:ETH:PORT |
Ethernet MAC |
74-69-69-2D-30-00 |
SYST:COMM:ETH:MAC |
Serial data bits |
8 |
– |
Serial stop bits |
1 |
– |
Serial port speed |
115200 |
– |
Serial port parity bit |
NONE |
– |
NTP service |
OFF |
SYST:COMM:ENAB 0, NTP |
NTP server |
europe.pool.ntp.org |
SYST:COMM:NTP |